Scope

1.1 This practice covers procedures for measuring and reporting the performance of Fourier-transform nuclear magnetic resonance spectrometers (FT-NMRs) using liquid samples.

1.2 This practice is not directly applicable to FT-NMR spectrometers outfitted to measure gaseous, anisotropically structured liquid, semi-solid, or solid samples; those set up to work with flowing sample streams; or those used to make hyperpolarization measurements.

1.3 This practice was expressly developed for FT-NMR spectrometers operating with proton resonance frequencies between 200 and 1200 MHz.

1.4 This practice is not directly applicable to continuous wave (scanning) NMR spectrometers.

1.5 This practice is not directly applicable to instruments using single-sideband detection.

1.6 Units—The values stated in SI units are to be regarded as the standard. No other units of measurement are included in this standard.

1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Significance and Use

4.1 This practice permits an analyst to compare the performance of an NMR spectrometer for a particular test on any given day with the instrument's prior performance for that test. The practice can also provide sufficient quantitative performance information for problem diagnosis and solving. If complete information about how a test is carried out is supplied and sufficient replicates are collected to substantiate statistical relevance, the tests in this practice can be used to establish the setting and meeting of relevant performance specifications. This practice is not necessarily meant for the comparison of different instruments with each other, even if the instruments are of the same type and model. This practice is not meant for the comparison of the performance of different instruments operated under conditions differing from those specified for a particular test.

Keywords

Fourier transform NMR; FT-NMR; gradient test; line shape test; NMR spectrometer; NMR standard; NMR test sample; performance test; probe temperature measurement; resolution test; sensitivity test;

To find similar documents by ASTM Volume:

03.06 (Molecular Spectroscopy; Surface Analysis)

To find similar documents by classification:

17.060 (Measurement of volume, mass, density, viscosity Including measuring instruments Volumetric measurements of petroleum products and natural gas, see 75.180.30)

This document comes with our free Notification Service, good for the life of the document.

This document is available in either Paper or PDF format.

Document Number

ASTM-E2977-15

Revision Level

2015 EDITION

Status

Superseded

Modification Type

Revision

Publication Date

June 1, 2015

Document Type

Practice

Page Count

30 pages

Committee Number

E13.15