ASTM-F3206 › Standard Guide for Assessing Medical Device Cytocompatibility with Delivered Cellular Therapies
The following bibliographic material is provided to assist you with your purchasing decision:
Scope
1.1 This guide outlines the parameters to consider when designing in vitro tests to assess the potential impact of a delivery device on a cellular product being dispensed. This guide does not provide specific protocols, but rather suggests what should be considered the minimum characterization necessary to assess device cytocompatibility. Topics discussed include selecting an appropriate cell line(s), cell physiology parameters to measure, and relevant test procedure variables. Only cells suspended in liquid and infused through a device are considered. Cell therapies paired with scaffolds, suspended in hydrogels, or administered via other methods (e.g., tissue grafting) are not included in the scope of this document. This document does not address physical characterization of delivery devices, such as mechanics, composition, or degradation.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Significance and Use
4.1 This guide is designed to assist medical device manufacturers as they develop new devices or qualify existing devices (e.g., catheters, needles) for delivering clinical cell therapies. Cytocompatibility considers the impact of the delivery device on the cells passing through the device during the delivery procedure. The biological safety of the device (e.g., the device’s cytotoxicity) should be addressed via other methods, such as ISO 10993-5. It is understood that this guide does not address testing of specific cellular products with specific delivery devices. Such testing may be required by regulatory authorities prior to clinical trial of cellular product or marketing applications. This guide outlines considerations to make the product qualification procedures more likely to succeed and more cost effective.
4.2 The key aspects of assessing device cytocompatibility include selecting a test cell line or cell lines and determining the cell physiology parameters that will be measured to make a determination of cytocompatibility. Acceptance criteria for designating a device as cytocompatible are not detailed here. It will be up to the delivery device end user to determine if the results of a cytocompatibility assessment are sufficient to consider that device cytocompatible. Delivery device lot to lot variability may impact cytocompatibility, therefore validated manufacturing processes should be considered when producing devices for cytocompatibility assessments.
Keywords
catheter; cell compatibility; cell therapy; cytocompatibility; medical device; stem cells;; ICS Number Code 11.100.01 (Laboratory medicine in general)
To find similar documents by ASTM Volume:
13.02 (Emergency Medical Services; Search and Rescue; Anesthetic and Respiratory Equipment)
To find similar documents by classification:
11.100.01 (Laboratory medicine in general)
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-F3206-17
Revision Level
2017 EDITION
Status
Current
Modification Type
New
Publication Date
June 15, 2017
Document Type
Guide
Page Count
4 pages
Committee Number
F04.43