ASTM-F3172 › Standard Guide for Design Verification Device Size and Sample Size Selection for Endovascular Devices
The following bibliographic material is provided to assist you with your purchasing decision:
Scope
1.1 This guide provides guidance for selecting an appropriate device size(s) and determining an appropriate sample size(s) (that is, number of samples) for design verification testing of endovascular devices. A methodology is presented to determine which device size(s) should be selected for testing to verify the device design adequately for each design input requirement (that is, test characteristic). Additionally, different statistical approaches are presented and discussed to help guide the developer to determine and justify sample size(s) for the design input requirement being verified. Alternate methodologies for determining device size selection and sample size selection may be acceptable for design verification.
1.2 This guide applies to physical design verification testing. This guide addresses in-vitro testing; in-vivo/animal studies are outside the scope of this guide. This guide does not directly address design validation; however, the methodologies presented may be applicable to in-vitro design validation testing. Guidance for sampling related to computational simulation (for example, sensitivity analysis and tolerance analysis) is not provided. Guidance for using models, such as design of experiments (DOE), for design verification testing is not provided. This guide does not address sampling across multiple manufacturing lots as this is typically done as process validation. Special considerations are to be given to certain tests such as fatigue (see Practice E739) and shelf life testing (see Section 8).
1.3 Regulatory guidance may exist for endovascular devices that should be considered for design verification device size and sample size selection.
1.4 Units—The values stated in SI units are to be regarded as the standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Significance and Use
4.1 The purpose of this guide is to provide guidance for selecting appropriate device size(s) and determining appropriate sample size(s) for design verification of endovascular devices. The device size(s) and sample size(s) for each design input requirement should be determined before testing. The device size(s) selected for verification testing should establish that the entire device matrix is able to achieve the design input requirements. If testing is not performed on all device sizes, justification should be provided.
4.2 The sample size justification and statistical procedures used to analyze the data should be based on sound scientific principles and should be suitable for reaching a justifiable conclusion. Insufficient sample size may lead to erroneous conclusions more often than desired.
4.3 Guidance regarding methodologies for determining device size selection and appropriate sample size is provided in Sections 5 and 6.
Keywords
attribute; normality; sample size; sampling plan; statistics; variables; verification;
To find similar documents by ASTM Volume:
13.02 (Emergency Medical Services; Search and Rescue; Anesthetic and Respiratory Equipment)
To find similar documents by classification:
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Customers who bought this document also bought:
ASTM-D4169Standard Practice for Performance Testing of Shipping Containers and Systems
ASTM-F88
Standard Test Method for Seal Strength of Flexible Barrier Materials
AAMI-ST72
Bacterial endotoxin - Test methods, routine monitoring and alternatives to batch testing
Document Number
ASTM-F3172-15R21
Revision Level
2015 R21 EDITION
Status
Current
Modification Type
New
Publication Date
Aug. 10, 2021
Document Type
Guide
Page Count
14 pages
Committee Number
F04.30