Scope
1.1 This test method covers a procedure for the fatigue testing of metallic tibial trays used in partial knee joint replacements.
1.2 This test method covers the procedures for the performance of fatigue tests on metallic tibial components using a cyclic, constant-amplitude force. It applies to tibial trays which cover either the medial or the lateral plateau of the tibia.
1.3 This test method may require modifications to accommodate other tibial tray designs.
1.4 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic tibial trays with unsupported mid-section of the condyle.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Significance and Use
4.1 This test method can be used to describe the effects of materials, manufacturing, and design variables on the fatigue performance of metallic tibial trays subject to cyclic loading for relatively large numbers of cycles.
4.2 The loading of tibial tray designs in vivo will, in general, differ from the loading defined in this practice. The results obtained here cannot be used to directly predict in vivo performance. However, this practice is designed to allow for comparisons between the fatigue performance of different metallic tibial tray designs, when tested under similar conditions.
4.3 In order for fatigue data on tibial trays to be comparable, reproducible, and capable of being correlated among laboratories, it is essential that uniform procedures be established.
Keywords
arthroplasty; orthopaedic medical devices; tibial components; unicondylar knee arthroplasty;
To find similar documents by ASTM Volume:
13.02 (Emergency Medical Services; Search and Rescue; Anesthetic and Respiratory Equipment)
To find similar documents by classification:
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Customers who bought this document also bought:
ASTM-E18Standard Test Methods for Rockwell Hardness of Metallic Materials
ASTM-A479
Standard Specification for Stainless Steel Bars and Shapes for Use in Boilers and Other Pressure Vessels
ASTM-A484
Standard Specification for General Requirements for Stainless Steel Bars, Billets, and Forgings
Document Number
ASTM-F3140-23
Revision Level
2023 EDITION
Status
Current
Modification Type
Revision
Publication Date
Sept. 12, 2023
Document Type
Test Method
Page Count
6 pages
Committee Number
F04.22