ASTM-F2132 › Standard Specification for Puncture Resistance of Materials Used in Containers for Discarded Medical Needles and Other Sharps (Withdrawn 2017)
The following bibliographic material is provided to assist you with your purchasing decision:
This specification provides the performance requirements and associated test procedures for the puncture resistance of materials used in the construction of containers for discarded medical needles and other sharps. It shall establish the average and minimum value of puncture force that the container materials must withstand under controlled laboratory conditions and normal room temperature when following the test procedure described herein. Materials meeting the specified performance requirements will be considered puncture resistant, but this specification will not evaluate the construction of, or provide pass/fail criteria for, sharps containers.
Scope
1.1 The purpose of this specification is to provide a test procedure and performance requirement for the puncture resistance of materials used in the construction of containers for discarded medical needles and other sharps. This test specification will establish (1) the average puncture force and (2) a minimum value of puncture force that container material(s) must withstand when following the test procedure described in Section 6. This specification shall be applicable to regions of uniform material and thickness, and needle contact areas as defined in 3.1.7 and 3.1.9. Materials meeting the performance requirements of Section 4 shall be considered “puncture-resistant.” This specification does not evaluate the construction of, or provide pass/fail criteria for, a sharps container.
1.2 This specification provides a test procedure to determine if all regions of one container meet the material puncture resistance requirements. It does not define the number of additional test containers required to achieve a statistically valid sample of a manufacturing lot or process. An appropriate sampling plan shall be determined by the test requester, as this depends upon the manufacturing process variability, manufacturing lot size, and other factors, such as end-user requirements.
1.3 This specification is intended to evaluate the performance of materials used in the construction or manufacture of sharps containers under controlled laboratory conditions, and at normal room temperature (see 6.1). (Warning—This specification only characterizes material puncture resistance at normal room temperatures. Applications of sharps containers outside the range of 23 ± 2°C (such as usage in emergency vehicles) require further material characterization by the product specifier to determine suitable use.)
1.4 The values stated in inch/pound are to be regarded as the standard. The SI values given in parentheses are for information only.
1.5 The following hazard caveat pertains only to the test procedure portion, Section 6, of this specification.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Keywords
material puncture resistance; needle disposal; puncture resistance; puncture-resistance testing; safety devices; sharps container; sharps disposal; Containers--environmental applications; Discarded medical instruments/supplies; Needles; Puncture testing; Sharps disposal; Surgical instruments--specifications; ICS Number Code 11.140 (Hospital equipment); 19.060 (Mechanical testing)
To find similar documents by ASTM Volume:
13.01 (Medical and Surgical Materials and Devices)
To find similar documents by classification:
19.060 (Mechanical testing Including testing equipment Mechanical testing of metals, see 77.040.10)
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Customers who bought this document also bought:
ISO-15223-1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
ASTM-D4169
Standard Practice for Performance Testing of Shipping Containers and Systems
ISO-10993-1
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Document Number
ASTM-F2132-01(2008)e1
Revision Level
2001 R08(E1) EDITION
Status
Superseded
Modification Type
Withdrawn
Publication Date
June 1, 2008
Document Type
Specification
Page Count
5 pages
Committee Number
F04.33