ASTM-E3029 › Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers
The following bibliographic material is provided to assist you with your purchasing decision:
Scope
1.1 This practice (1)2 describes three methods for determining the relative spectral correction factors for grating-based fluorescence spectrometers in the ultraviolet-visible spectral range. These methods are intended for instruments with a 0°/90° transmitting sample geometry. Each method uses different types of transfer standards, including 1) a calibrated light source (CS), 2) a calibrated detector (CD) and a calibrated diffuse reflector (CR), and 3) certified reference materials (CRMs). The wavelength region covered by the different methods ranges from 250 to 830 nm with some methods having a broader range than others. Extending these methods to the near infrared (NIR) beyond 830 nm will be discussed briefly, where appropriate. These methods were designed for scanning fluorescence spectrometers with a single channel detector, but can also be used with a multichannel detector, such as a diode array or a CCD.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Significance and Use
3.1 Calibration of the responsivity of the detection system for emission (EM) as a function of EM wavelength (λEM), also referred to as spectral correction of emission, is necessary for successful quantification when intensity ratios at different EM wavelengths are being compared or when the true shape or peak maximum position of an EM spectrum needs to be known. Such calibration methods are given here and summarized in Table 1. This type of calibration is necessary because the spectral responsivity of a detection system can change significantly over its useful wavelength range (see Fig. 1). It is highly recommended that the wavelength accuracy (see Test Method E388) and the linear range of the detection system (see Guide E2719 and Test Method E578) be determined before spectral calibration is performed and that appropriate steps are taken to insure that all measured intensities during this calibration are within the linear range. For example, when using wide slit widths in the monochromators, attenuators may be needed to attenuate the excitation beam or emission, thereby, decreasing the fluorescence intensity at the detector. Also note that when using an EM polarizer, the spectral correction for emission is dependent on the polarizer setting. (2) It is important to use the same instrument settings for all of the calibration procedures mentioned here, as well as for subsequent sample measurements.
3.2 When using CCD or diode array detectors with a spectrometer for λEM selection, the spectral correction factors are dependent on the grating position of the spectrometer. Therefore, the spectral correction profile versus λEM must be determined separately for each grating position used. (3)
3.3 Instrument manufacturers often provide an automated procedure and calculation for a spectral correction function for emission, or they may supply a correction that was determined at the factory. This correction can often be applied during spectral collection or as a post-collection correction. The user should be advised to verify that the automated vendor procedure and calculation or supplied correction are performed and determined according to the guidelines given within this standard.
Keywords
calibration; fluorescence; fluorometer; luminescence; qualification; reference materials; spectral correction; spectrometer; spectroscopy; validation;
To find similar documents by ASTM Volume:
03.06 (Molecular Spectroscopy; Surface Analysis)
To find similar documents by classification:
17.180.01 (Optics and optical measurements in general)
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-E3029-15R23
Revision Level
2015 R23 EDITION
Status
Current
Modification Type
New
Publication Date
Jan. 6, 2023
Document Type
Practice
Page Count
5 pages
Committee Number
E13.01