Standard Practice for Blood Irradiation Dosimetry
ORDER

Price:

$76.00        


Want this as a site license?

Changes from the previous issue

A redline edition is available for this document, with all changes visible. Ask Document Center Inc. for more information.

Scope

1.1 This practice outlines the irradiator installation qualification program and the dosimetric procedures to be followed during operational qualification and performance qualification of the irradiator. Procedures for the routine radiation processing of blood product (blood and blood components) are also given. If followed, these procedures will help ensure that blood product exposed to gamma radiation or X-radiation (bremsstrahlung) will receive absorbed doses with a specified range.

1.2 This practice covers dosimetry for the irradiation of blood product for self-contained irradiators (free-standing irradiators) utilizing radionuclides such as 137Cs and   60Co, or X-radiation (bremsstrahlung). The absorbed dose range for blood irradiation is typically 15 Gy to 50 Gy.

1.3 The photon energy range of X-radiation used for blood irradiation is typically from 40 keV to 300 keV.

1.4 This practice also covers the use of radiation-sensitive indicators for the visual and qualitative indication that the product has been irradiated (see ISO/ASTM Guide 51539).

1.5 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing and describes a means of achieving compliance with the requirements of ISO/ASTM Practice 52628 for dosimetry performed for blood irradiation. It is intended to be read in conjunction with ISO/ASTM Practice 52628.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability or regulatory limitations prior to use.

Significance and Use

4.1 Blood and blood components are irradiated to predetermined absorbed doses to inactivate viable lymphocytes to help prevent transfusion-induced graft-versus-host disease (GVHD) in certain immunocompromised patients and those receiving related-donor products (1, 2).9

4.2 The assurance that blood and blood components have been properly irradiated is of crucial importance for patient health. This shall be demonstrated by means of accurate absorbed-dose measurements on the product, or in simulated product.

4.3 Blood and blood components are usually irradiated using gamma radiation from 137Cs or 60Co sources, or X-radiation from X-ray units.

4.4 Blood irradiation specifications include a lower limit of absorbed dose, and may include an upper limit or central target dose. For a given application, any of these values may be prescribed by regulations that have been established on the basis of available scientific data (see 2.6).

4.5 For each blood irradiator, an absorbed-dose rate at a reference position within the canister is measured as part of irradiator acceptance testing using a reference-standard dosimetry system. That reference-standard measurement is used to establish operating parameters so as to deliver specified dose to blood and blood components.

4.6 Absorbed-dose measurements are performed within the blood or blood-equivalent volume for determining the absorbed-dose distribution. Such measurements are often performed using simulated product (for example, polystyrene is considered blood equivalent for 137Cs photon energies).

4.7 Dosimetry is part of a measurement management system that is applied to ensure that the radiation process meets predetermined specifications (see ISO/ASTM Practice 52628).

4.8 Blood and blood components are usually irradiated in chilled or frozen condition. Care should be taken, therefore, to ensure that the dosimeters and radiation-sensitive indicators can be used under such temperature conditions.

4.9 Proper documentation and record keeping are critical components of a radiation process. Documentation and record keeping requirements may be specified by regulatory authorities or may be given in the corporation’s quality policy.

4.10 Response of most dosimeters has significant energy dependence at photon energies of less than 100 keV, so proper care must be exercised when measuring absorbed dose in that energy range.

Keywords

absorbed dose; absorbed-dose mapping; blood; blood components; blood products; blood irradiation; bremsstrahlung; dosimeter; dosimetry system; gamma radiation; irradiator; ionizing radiation; measurement quality assurance plan; measurement uncertainty; radiation-sensitive indicators; reference-standard dosimeter; routine dosimeter; transfer-standard dosimeter; ICS 17.240; X-radiation;; ICS Number Code 11.040.20 (Transfusion, Infusion); 17.240 (Radiation measurements)

To find similar documents by ASTM Volume:

12.02 (Nuclear (II), Solar, and Geothermal Energy)

To find similar documents by classification:

11.040.20 (Transfusion, infusion and injection equipment Including blood packs Syringes, needles and catheters, see 11.040.25)

11.100 (Laboratory medicine)

17.240 (Radiation measurements Including dosimetry Radiation protection, see 13.280)

This document comes with our free Notification Service, good for the life of the document.

This document is available in either Paper or PDF format.

Document Number

ASTM-E1939-17

Revision Level

2017 EDITION

Status

Current

Modification Type

Revision

Publication Date

Feb. 15, 2017

Document Type

Practice

Page Count

13 pages

Committee Number

E61.04