ASTM-C1458 › Standard Test Method for Nondestructive Assay of Plutonium, Tritium and 241Am by Calorimetric Assay
The following bibliographic material is provided to assist you with your purchasing decision:
Changes from the previous issue
A redline edition is available for this document, with all changes visible. Ask Document Center Inc. for more information.
Scope
1.1 This test method describes the nondestructive assay (NDA) of plutonium, tritium, and 241Am using heat flow calorimetry. For plutonium the typical range of applicability, depending on the isotopic composition, corresponds to ~0.1 g to ~5 g quantities while for tritium the typical range extends from ~0.001 g to ~400 g. This test method can be applied to materials in a wide range of container sizes up to 380 L. It has been used routinely to assay items whose thermal power ranges from 0.001 W to 135 W.
1.2 This test method requires knowledge of the relative abundances of the plutonium isotopes and the 241Am/Pu mass ratio to determine the total plutonium mass.
1.3 This test method provides a direct measure of tritium content.
1.4 This test method provides a measure of 241Am either as a single isotope or mixed with plutonium.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Significance and Use
5.1 This test method is considered to be the most accurate NDA technique for the assay of many physical forms of Pu. Isotopic measurements by gamma-ray spectroscopy or destructive analysis techniques are part of this test method when it is applied to the assay of Pu.
5.1.1 Calorimetry has been applied to a wide variety of Pu-bearing solids including metals, alloys, oxides, fluorides, mixed Pu-U oxides, mixed oxide fuel pins, waste, and scrap, for example, ash, ash heels, salts, crucibles, and graphite scarfings) (2, 3). This test method has been routinely used at U.S. and European facilities for Pu process measurements and nuclear material accountability since the mid 1960’s (2-9).
5.1.2 Pu-bearing materials have been measured in calorimeter containers ranging in size from about 0.025 m to about 0.63 m in diameter and from about 0.076 m to about 1.38 m in height.
5.1.3 Gamma-ray spectroscopy typically is used to determine the Pu isotopic composition and 241Am to Pu ratio (see Test Method C1030). However, isotopic information from mass spectrometry and alpha counting measurements may be used instead (see Test Method C697).
5.2 This test method is considered to be the most accurate NDA method for the measurement of tritium. For many physical forms of tritium compounds calorimetry is currently the only practical measurement technique available.
5.3 Physical standards representative of the materials being assayed are not required for the test method.
5.3.1 This test method is largely independent of the elemental distribution of the nuclear materials in the matrix.
5.3.2 The accuracy of the method can be degraded for materials with inhomogeneous isotopic composition.
5.4 The thermal power measurement is traceable to national measurement systems through electrical standards used to directly calibrate the calorimeters or to calibrate secondary 238Pu heat standards.
5.5 Heat-flow calorimetry has been used to prepare secondary standards for neutron and gamma-ray assay systems (7-12).
5.6 Four parameters of the item and the item packaging affect measurement time. These four parameters are density, mass, thermal conductivity, and change in temperature. The measurement well of passive calorimeters will also affect measurement time because it too will need to come to the new equilibrium temperature. Calorimeters operated in power compensation mode maintain a constant measurement well temperature and have no additional effect on measurement time.
5.6.1 Calorimeter measurement times range from 20 minutes (13) for smaller, temperature-conditioned containers up to 72 h (14) for larger containers and items with long thermal-time constants.
5.6.2 Measurement times may be reduced by using equilibrium prediction techniques, by temperature preconditioning of the item to be measured, by operating the calorimeter using the power compensation technique, or by optimization of the item container (low thermal mass and high thermal conductivity) and packaging.
Keywords
accountability; calorimetry; inventory; nondestructive assay; plutonium; quantification; tritium;; ICS Number Code 27.120.30 (Fissile materials and nuclear fuel technology)
To find similar documents by ASTM Volume:
To find similar documents by classification:
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-C1458-16
Revision Level
2016 EDITION
Status
Current
Modification Type
Revision
Publication Date
April 1, 2016
Document Type
Test Method
Page Count
13 pages
Committee Number
C26.10